© Adam Equipment Company 2013 Adam Equipment HCB J Precious Metals and Density Balance (P.N. 308660002 Revision D1, June 2013)
8 © Adam Equipment Company 2013 3) Remove the shipping protection screw and transit disc from the top of the balance. 4)
9 © Adam Equipment Company 2013 5) Place the liquid chamber onto the support. Ensure the line marking the water level is to the fr
10 © Adam Equipment Company 2013 7) Position the sample pan in the beaker area and connect the 4 suspension wires to the rings in th
11 © Adam Equipment Company 2013 When you are preparing to do a Density Test it will be necessary to fill the chamber
12 © Adam Equipment Company 2013 2.4 KEYPAD / DISPLAY FUNCTION The keypad and display have a number of features as shown and explain
13 © Adam Equipment Company 2013 2.5 REAR PANEL The rear panel has connectors for the power, RS232, remote display and te
14 © Adam Equipment Company 2013 3.0 LOCATING AND PROTECTING YOUR BALANCE In order to keep your balance functioning at its
15 © Adam Equipment Company 2013 4.0 BASIC OPERATION 4.1 TURNING ON THE BALANCE Plug in the unit using the adapter or use
16 © Adam Equipment Company 2013 5.0 DENSITY FUNCTION Once the HCB J is prepared for a density test and the parameters have been set
17 © Adam Equipment Company 2013 display will show “HOLD”. If the Hold function is set to automatic you are not require
© Adam Equipment Company 2013
18 © Adam Equipment Company 2013 table below), the karat value and content value will not be displayed. An error message
19 © Adam Equipment Company 2013 6.0 DENSITY PARAMETERS To use the density function it is necessary to set parameters desc
20 © Adam Equipment Company 2013 temperature of the water, selections are either using the automatic temperature sensor or
21 © Adam Equipment Company 2013 options are Silver (Ag), Copper (Cu), Silver: Copper in a 1:1 ratio (AG:CU 1-1), Silver: Copper in a
22 © Adam Equipment Company 2013 When the sample is Platinum mixed with an alloy the user needs to select the option that describes t
23 © Adam Equipment Company 2013 7.0 PARAMETERS The balance has 8 parameters that can be set by the user. FUNCTION SECTION DESCRI
24 © Adam Equipment Company 2013 7.1 ENABLING WEIGHING UNITS, F1F1F1F1 UUUUNNNNTTTT You can enable and disable the weighing units a
25 © Adam Equipment Company 2013 The following table shows different units which are available to the user and the conv
26 © Adam Equipment Company 2013 7.2 HOLD FUNCTION, F2F2F2F2 HHHHLLLLDDDD The hold function may be set to automatic, manual
27 © Adam Equipment Company 2013 7.3 BACKLIGHT, F3F3F3F3 bbbbLLLL The backlight may be enabled or disabled by the user. If th
1 © Adam Equipment Company 2013 CONTENTS 1.0 INTRODUCTION ...
28 © Adam Equipment Company 2013 5) The following options are available for setting the serial output functions: Mode Print Feature
29 © Adam Equipment Company 2013 7.5 AUTO POWER OFF, F5F5F5F5 OFOFOFOFFFFF The auto power off function helps conserve
30 © Adam Equipment Company 2013 7.7 ADJUST THE VALUE OF THE INTERNAL MASS, F7F7F7F7 CCCCAAAA The internal mass value stored
31 © Adam Equipment Company 2013 7.8 USER TECHNICAL PARAMETERS, F8F8F8F8 SSSSEtEtEtEt The balance has a number of technical param
32 © Adam Equipment Company 2013 8.0 USER CALIBRATION The Highland series of balances comes standard with HandiCalTM internal calibr
33 © Adam Equipment Company 2013 External Calibration 1) Press and hold the [Air/Liquid] and [Density] keys at the same t
34 © Adam Equipment Company 2013 9.0 SERIAL INTERFACE SPECIFICATIONS The balance comes equipped with an RS-232 Serial Interface. The
35 © Adam Equipment Company 2013 Format for simple weight only output: Input command format: The balance can be controlled with
36 © Adam Equipment Company 2013 10.0 REMOTE DISPLAY The remote display plugs into the 6 pin circular connector on the r
37 © Adam Equipment Company 2013 12.0 TROUBLE SHOOTING 12.1 ERROR MESSAGES If an error message is shown, repeat the step that cause
2 © Adam Equipment Company 2013 1.0 INTRODUCTION Thank you for purchasing your new HCB J Precious Metals and Density Balance. As wi
38 © Adam Equipment Company 2013 12.2 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories contac
39 © Adam Equipment Company 2013 13.0 SPECIFICATIONS 13.1 TECHNICAL SPECIFICATIONS Model # HCB J123 HCB J302 HCB J1002 HCB J2
40 © Adam Equipment Company 2013 14.0 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for the com
© Adam Equipment Company 2013 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised
© Adam Equipment Company 2013 ADAM EQUIPMENT is an ISO 9001:2008 certified global company with more than 35 years experience in the production
3 © Adam Equipment Company 2013 2) Remove the shipping protection screw and transit disc from the top of the balance 3)
4 © Adam Equipment Company 2013 5) Replace the beaker onto the beaker support and pull the Temperature Sensor cable through the cl
5 © Adam Equipment Company 2013 7) Plug the Temperature sensor into the 5 pin connector, and the remote Indicator into the 6 pin co
6 © Adam Equipment Company 2013 2.2 HCB J302, J1002 AND J2002 SETTING UP YOUR BALANCE 1) Remove the balance and all accessories f
7 © Adam Equipment Company 2013