HCB J Precious Metals and Density
Balance
(P.N. 308660002 Revision D1, June 2013)
Software 4.2 and above
Page view 0
1 2 3 4 5 6 ... 43 44

Summary of Contents

Page 1 - Adam Equipment

© Adam Equipment Company 2013 Adam Equipment HCB J Precious Metals and Density Balance (P.N. 308660002 Revision D1, June 2013)

Page 2

8 © Adam Equipment Company 2013 3) Remove the shipping protection screw and transit disc from the top of the balance. 4)

Page 3 - CONTENTS

9 © Adam Equipment Company 2013 5) Place the liquid chamber onto the support. Ensure the line marking the water level is to the fr

Page 4 - 2.0 SET

10 © Adam Equipment Company 2013 7) Position the sample pan in the beaker area and connect the 4 suspension wires to the rings in th

Page 5

11 © Adam Equipment Company 2013 When you are preparing to do a Density Test it will be necessary to fill the chamber

Page 6

12 © Adam Equipment Company 2013 2.4 KEYPAD / DISPLAY FUNCTION The keypad and display have a number of features as shown and explain

Page 7

13 © Adam Equipment Company 2013 2.5 REAR PANEL The rear panel has connectors for the power, RS232, remote display and te

Page 8

14 © Adam Equipment Company 2013 3.0 LOCATING AND PROTECTING YOUR BALANCE In order to keep your balance functioning at its

Page 9

15 © Adam Equipment Company 2013 4.0 BASIC OPERATION 4.1 TURNING ON THE BALANCE Plug in the unit using the adapter or use

Page 10 -

16 © Adam Equipment Company 2013 5.0 DENSITY FUNCTION Once the HCB J is prepared for a density test and the parameters have been set

Page 11

17 © Adam Equipment Company 2013 display will show “HOLD”. If the Hold function is set to automatic you are not require

Page 12

© Adam Equipment Company 2013

Page 13

18 © Adam Equipment Company 2013 table below), the karat value and content value will not be displayed. An error message

Page 14

19 © Adam Equipment Company 2013 6.0 DENSITY PARAMETERS To use the density function it is necessary to set parameters desc

Page 15

20 © Adam Equipment Company 2013 temperature of the water, selections are either using the automatic temperature sensor or

Page 16 - BALANCE

21 © Adam Equipment Company 2013 options are Silver (Ag), Copper (Cu), Silver: Copper in a 1:1 ratio (AG:CU 1-1), Silver: Copper in a

Page 17 - OPERATION

22 © Adam Equipment Company 2013 When the sample is Platinum mixed with an alloy the user needs to select the option that describes t

Page 18 - FUNCTION

23 © Adam Equipment Company 2013 7.0 PARAMETERS The balance has 8 parameters that can be set by the user. FUNCTION SECTION DESCRI

Page 19

24 © Adam Equipment Company 2013 7.1 ENABLING WEIGHING UNITS, F1F1F1F1 UUUUNNNNTTTT You can enable and disable the weighing units a

Page 20

25 © Adam Equipment Company 2013 The following table shows different units which are available to the user and the conv

Page 21 - PARAMETERS

26 © Adam Equipment Company 2013 7.2 HOLD FUNCTION, F2F2F2F2 HHHHLLLLDDDD The hold function may be set to automatic, manual

Page 22

27 © Adam Equipment Company 2013 7.3 BACKLIGHT, F3F3F3F3 bbbbLLLL The backlight may be enabled or disabled by the user. If th

Page 23

1 © Adam Equipment Company 2013 CONTENTS 1.0 INTRODUCTION ...

Page 24

28 © Adam Equipment Company 2013 5) The following options are available for setting the serial output functions: Mode Print Feature

Page 25 - 7.0 PARAMETERS

29 © Adam Equipment Company 2013 7.5 AUTO POWER OFF, F5F5F5F5 OFOFOFOFFFFF The auto power off function helps conserve

Page 26

30 © Adam Equipment Company 2013 7.7 ADJUST THE VALUE OF THE INTERNAL MASS, F7F7F7F7 CCCCAAAA The internal mass value stored

Page 27

31 © Adam Equipment Company 2013 7.8 USER TECHNICAL PARAMETERS, F8F8F8F8 SSSSEtEtEtEt The balance has a number of technical param

Page 28

32 © Adam Equipment Company 2013 8.0 USER CALIBRATION The Highland series of balances comes standard with HandiCalTM internal calibr

Page 29

33 © Adam Equipment Company 2013 External Calibration 1) Press and hold the [Air/Liquid] and [Density] keys at the same t

Page 30 - Mode Print Feature

34 © Adam Equipment Company 2013 9.0 SERIAL INTERFACE SPECIFICATIONS The balance comes equipped with an RS-232 Serial Interface. The

Page 31

35 © Adam Equipment Company 2013 Format for simple weight only output: Input command format: The balance can be controlled with

Page 32

36 © Adam Equipment Company 2013 10.0 REMOTE DISPLAY The remote display plugs into the 6 pin circular connector on the r

Page 33

37 © Adam Equipment Company 2013 12.0 TROUBLE SHOOTING 12.1 ERROR MESSAGES If an error message is shown, repeat the step that cause

Page 34 - CALIBRATION

2 © Adam Equipment Company 2013 1.0 INTRODUCTION Thank you for purchasing your new HCB J Precious Metals and Density Balance. As wi

Page 35

38 © Adam Equipment Company 2013 12.2 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories contac

Page 36 - SPECIFICATIONS

39 © Adam Equipment Company 2013 13.0 SPECIFICATIONS 13.1 TECHNICAL SPECIFICATIONS Model # HCB J123 HCB J302 HCB J1002 HCB J2

Page 37

40 © Adam Equipment Company 2013 14.0 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for the com

Page 38 - 11.0 TEMPERATURE

© Adam Equipment Company 2013 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised

Page 39 - SHOOTING

© Adam Equipment Company 2013 ADAM EQUIPMENT is an ISO 9001:2008 certified global company with more than 35 years experience in the production

Page 40

3 © Adam Equipment Company 2013 2) Remove the shipping protection screw and transit disc from the top of the balance 3)

Page 41 - 13.0 SPECIFICATIONS

4 © Adam Equipment Company 2013 5) Replace the beaker onto the beaker support and pull the Temperature Sensor cable through the cl

Page 42

5 © Adam Equipment Company 2013 7) Plug the Temperature sensor into the 5 pin connector, and the remote Indicator into the 6 pin co

Page 43

6 © Adam Equipment Company 2013 2.2 HCB J302, J1002 AND J2002 SETTING UP YOUR BALANCE 1) Remove the balance and all accessories f

Page 44 - ADAM EQUIPMENT

7 © Adam Equipment Company 2013